Device product
Syngo.Via VB20A model 1049610
Z-1778-2018
Product summary
- Event
- Event 79888
- Status
- Terminated
- Classification
- Class II
- Quantity
- N/A
- Official record key
device-enforcement:Z-1778-2018
Official wording
Reason: Functionality in the report sections "Findings Information" and "Summary of Measured Findings" not functioning properly. The corrected values modified in the report are neither saved, nor printed, nor sent to the information system. The system will still use original values.
Code information: All
Distribution pattern: Nationwide
Derived failure modes
-
Unknown
Functionality in the report sections "Findings Information" and "Summary of Measured Findings" not functioning properly. The corrected values modified in the report are neither saved, nor printed, nor sent to the information system. The system will still use original values.