Device product

lmmunalysis EDDP Urine Control Set 1 The EDDP Urine Controls (High 125 ng/ml and Low 75 ng/ml) are used as control materials in the ammonolysis EDDP Specific Urine Enzyme Immunoassay, which has a cutoff of 100 ng/ml. This test provides only a preliminary analytical test result and a more specific alternate chemical method, such as Gas Chromatography/Mass Spectrometry or Liquid Chromatography/Tandem Mass Spectrometry, must be used to obtain a confirmed analytical result.

Z-2140-2018

November 28, 2017

Class III

Product summary

Firm
Alere San Diego Inc. DBA Immunalysis Corporation
Event
Event 80096
Status
Terminated
Classification
Class III
Quantity
37 kits
Official record key
device-enforcement:Z-2140-2018

Official wording

Reason: The High (125 ng/ml) Control does not meet specification - it has been confirmed to produce values below 100 ng/mL when run on validated clinical chemistry analyzers, which is below the 100 ng/mL cutoff of the EDDP Specific Urine Enzyme Immunoassay with which this control is utilized.

Code information: Part Number: 3006-5, Lot Number: EK16135

Distribution pattern: US Distribution to the states of : CA, CO, CT, FL, LA, MD, NC, NJ, OH, OK, TN, TX, WI, and WV

Derived failure modes

  • Potency or specification failure

    reason.potency_specification_failure · v1.0.0

    does not meet specification