Device product
Femoral Nail, RIGHT T2 GTN ¿9x320 mm, Product Number 18510932S
Z-2405-2018
Product summary
- Firm
- Stryker GmbH
- Event
- Event 80131
- Status
- Terminated
- Classification
- Class II
- Quantity
- Unavailable
- Official record key
device-enforcement:Z-2405-2018
Official wording
Reason: The manufacturer has discovered that potentially out-of-specification products may have left the factory. Non-conforming cannulation of the nails may result in reduced component strength and potentially premature nail breakage.
Code information: GTIN: 04546540647320; Lot code: K076A80
Distribution pattern: The products were distributed to the following US states: AL, Co, FL, KY, NC, TN, UT, and WA.
Derived failure modes
-
Potency or specification failure
out-of-specification