Device product
ARTIS Pheno Interventional Fluoroscopic X-Ray System, Model Number 10849000
Z-3190-2018
Product summary
- Event
- Event 80905
- Status
- Terminated
- Classification
- Class II
- Quantity
- 429 total
- Official record key
device-enforcement:Z-3190-2018
Official wording
Reason: Temperature of the detector may fall out of its designed working range (cool down) due to incorrectly configured temperature control unit. This unintentional system behavior may occur while the unit is powered down for long periods (e.g. at night).
Code information: 164013 164011
Distribution pattern: US nationwide distribution.
Derived failure modes
-
Unknown