Device product
CombiDiagnost R90 (Stitching Patient Support) 709030
Z-2847-2018
Product summary
- Event
- Event 80577
- Status
- Terminated
- Classification
- Class II
- Quantity
- Unavailable
- Official record key
device-enforcement:Z-2847-2018
Official wording
Reason: The hook does not securely hold the footplate in the vertical position. Therefore, the footplate can fall or swing down and injury operator or patient.
Code information: Serial Numbers: SN17000130 SN17000018
Distribution pattern: US Nationwide; International to 66 countries
Derived failure modes
-
Unknown
The hook does not securely hold the footplate in the vertical position. Therefore, the footplate can fall or swing down and injury operator or patient.