AXIOM Artis with Laird Cooling Unit for SSFD: Material # 7555118. Artis is an angiography system developed for single and biplane diagnostic imaging and interventional procedures.

Z-2525-2018

May 25, 2018

Class II

Product summary

Firm
Siemens Medical Solutions USA, Inc
Event
Event 80476
Status
Terminated
Classification
Class II
Quantity
6 units US
Official record key
device-enforcement:Z-2525-2018

Official wording

Reason: Due to a defective sealing, coolant may inflow into the electrical parts of the system cabinet, which could cause a short circuit and cause the system to shut down suddenly during an ongoing procedure

Code information: Laird Cooling Unit for SSFD: Material # 7555118.