Device product

Discovery MI Digital Ready

Z-2579-2018

May 24, 2018

Class II

Product summary

Firm
GE Healthcare, LLC
Event
Event 80484
Status
Terminated
Classification
Class II
Quantity
42 units total
Official record key
device-enforcement:Z-2579-2018

Official wording

Reason: DIGITAL ReadyCT/PET systems recently installed with a GE supplied A! electrical panel may not be properly wired to the partial system UPS (Uninteruptible Power Supply).

Code information: Model Number 5986030. a. Serial Number PTDMI1800003PT, System ID Number RADNETPT1839, UDI Number 01008406821209751118010421PTDMI1800003PT; b. Serial Number PTDMI1800005PT, System ID Number RADNETPT1838, UDI Number 01008406821209751118010421PTDMI1800003PT; c. Serial Number PTDMI1800009PT, System ID Number 813745DMIDR2, UDI Number 01008406821209751118021921PTDMI1800009PT

Distribution pattern: Nationwide distribution to AL, CA, CO, FL, IL, IN, MA, MD, MI, MS, NC, NJ, NY,OH, OK, SC, TN, TX, VA.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    DIGITAL ReadyCT/PET systems recently installed with a GE supplied A! electrical panel may not be properly wired to the partial system UPS (Uninteruptible Power Supply).