Device product

Diamond Coated Tips and Burs are accessories to the air powered Bone drill (VARIOSURG with and without light) The intended use of the tips is based on the parent device. For the VarioSurg, in addition to the scaling functions, diamond coated tips are used for socket lift, sinus lift, and implant preparation.

Z-2673-2016

July 08, 2016

Class II

Product summary

Firm
Nakanishi Inc.
Event
Event 74666
Status
Terminated
Classification
Class II
Quantity
2574 units
Official record key
device-enforcement:Z-2673-2016

Official wording

Reason: Through an error, the labeling failed to include the symbol or other text indicating that the diamond coated tips were single use only. A contributing factor to this error is that in other markets such as Japan, the diamond coated tips are multiple use and are not limited to single use only. The tips are stainless steel with a coating of fine diamond powder. The tips are sold non-sterile and are to be sterilized prior to first use. The tips were cleared as single use only. This problem was discovered during an ongoing, proactive, quality system improvement plan that was initiated by Nakanishi in October 2014. Because the correct instructions for sterilization are not available, it is possible that the devices might not be sterile prior to use. The absence of the single use statement introduces a risk of cross-contamination if the tips are reprocessed and reused.

Code information: Waiting for clarification from the firm.

Distribution pattern: Nationwide Distribution to Illinois, Florida and Georgia.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Through an error, the labeling failed to include the symbol or other text indicating that the diamond coated tips were single use only. A contributing factor to this error is that in other markets such as Japan, the diamond coated tips are multiple use and are not limited to single use only. The tips are stainless steel with a coating of fine diamond powder. The tips are sold non-sterile and are to be sterilized prior to first use. The tips were cleared as single use only. This problem was discovered during an ongoing, proactive, quality system improvement plan that was initiated by Nakanishi in October 2014. Because the correct instructions for sterilization are not available, it is possible that the devices might not be sterile prior to use. The absence of the single use statement introduces a risk of cross-contamination if the tips are reprocessed and reused.