Device product
6" Smallbore Ext Set w/6-Port NanoClave Manifold, Check Valve, Clamp, Rotating Luer.
Z-2816-2018
Product summary
- Event
- Event 80460
- Status
- Terminated
- Classification
- Class II
- Quantity
- 300 pouches
- Official record key
device-enforcement:Z-2816-2018
Official wording
Reason: ICU Medical is expanding the scope of their previously initiated recall to include additional I.V. Administration Sets with NanoClave manifold due to the potential to develop a leak or for separation between the female luer and the NanoClave manifold.
Code information: Item Number AM6130. a. Lot Number 3414534, UDI Number (01) 1 0887709 08771 9 (17) 220201 (30) 50 (10) 3414534, Expiration Date 02/01/2022; b. Lot Number 3442461, UDI Number (01) 1 0887709 08771 9 (17) 220501 (30) 50 (10) 3442461, Expiration Date 05/01/2022; b. Lot Number 3509517, UDI Number (01) 1 0887709 08771 9 (17) 220801 (30) 50 (10) 3509517, Expiration Date 08/01/2022
Distribution pattern: US distribution to AZ, MA, NC, OH, TX, VA, WA. Worldwide distribution to Belgium, France, Germany.
Derived failure modes
-
Unknown
ICU Medical is expanding the scope of their previously initiated recall to include additional I.V. Administration Sets with NanoClave manifold due to the potential to develop a leak or for separation between the female luer and the NanoClave manifold.