Carina Mobile Lifts, Models: a) Carina350EM, REF 6060011 b) Carina350EML, REF 60600012 c) Carina350EE, REF 60600009

Z-0014-2018

July 28, 2016

Class II

Product summary

Firm
Handicare AB
Event
Event 78130
Status
Terminated
Classification
Class II
Quantity
9 units
Official record key
device-enforcement:Z-0014-2018

Official wording

Reason: A limited amount of Carina mobile lifts have been delivered with a too small locking pin. If a locking pin smaller than the correct size is mounted on the Carina mobile lift, the locking pin is likely to cause increased wear on the plastic bushings. In case of continued use, these components may cause wear on the metal of the piston rod eye.

Code information: a) Carina350EM, REF 6060011, Serial numbers: 661102000 - 661102160 b) Carina350EML, REF 60600012, Serial numbers: 661200500 - 661200584 c) Carina350EE, REF 60600009, Serial numbers: 660900500 -660900571

Distribution pattern: OH, PA, MD

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    A limited amount of Carina mobile lifts have been delivered with a too small locking pin. If a locking pin smaller than the correct size is mounted on the Carina mobile lift, the locking pin is likely to cause increased wear on the plastic bushings. In case of continued use, these components may cause wear on the metal of the piston rod eye.