Device product
Dimension Vista¿ Mass Creatine Kinase (MMB) Flex¿ reagent cartridge
Z-1672-2018
Product summary
- Event
- Event 79505
- Status
- Terminated
- Classification
- Class II
- Quantity
- N/A
- Official record key
device-enforcement:Z-1672-2018
Official wording
Reason: Concentrations for the level of biotin that does not interfere is not currently listed in the instruction for use for the Dimension Vista CTNI, Dimension Vista MMB, Dimension Sirolimus, ADVIA Centaur FOL, ADVIA Centaur aHBcM and the following IMMULITE assays, 3gAllergy Specific IgE, anti-HBc, BR-MA (CA15-3), CEA, Folic Acid, Gastrin, OM-MA (CA125), Thyroglobulin, and Vitamin B12, and these assays are susceptible to interference from biotin.
Code information: MMB, colorimetric method, cpk or isozymes, K6420, SMN# 10445097 Lot codes: 16284BD 16312BB 17011BD 17058BD 17093BC 17136BE 17186BE 17235BC 17268BA
Distribution pattern: Nationwide and Canada, Mexico, and Thailand
Derived failure modes
-
Unknown
Concentrations for the level of biotin that does not interfere is not currently listed in the instruction for use for the Dimension Vista CTNI, Dimension Vista MMB, Dimension Sirolimus, ADVIA Centaur FOL, ADVIA Centaur aHBcM and the following IMMULITE assays, 3gAllergy Specific IgE, anti-HBc, BR-MA (CA15-3), CEA, Folic Acid, Gastrin, OM-MA (CA125), Thyroglobulin, and Vitamin B12, and these assays are susceptible to interference from biotin.