Skip to content
Recall Observatory FDA recall evidence

Device product

SORIN GROUP, SMART PERFUSION PACK, NEONATE/PEDIATRIC PACK, REF 627127702, STERILE EO, Rx Only, 1 EA

Z-2792-2017

June 29, 2017

Class II

Product summary

Firm
Sorin Group USA, Inc.
Event
Event 77810
Status
Terminated
Classification
Class II
Quantity
54 units
Official record key
device-enforcement:Z-2792-2017

Official wording

Reason: Medtronic informed LivaNova that some of their DLP Pressure Disposable Display Sets are packaged in pouches which may have small pinholes (approximately 1mm or less in diameter). The result is a potential sterile barrier breach in the display sets. These display sets are included with LivaNova Perfusion Tube Sets (PTS) which were then distributed to customers.

Code information: Lot Numbers: 1620000113, 1622400049, 1624300090, 1631200072, 1632100059, 1632800059, 1633300054, 1705100057, 1711500053

Distribution pattern: Nationwide

Derived failure modes

  • Sterility assurance

    reason.sterility_assurance · v1.0.0

    sterile barrier breach