Device product
Pressure Injectable Multi-Lumen Central Venous Catheterization Kit with Blue FlexTip¿, ARROWg+ard Blue PLUS¿ Catheter, Sharps Safety Features and Maximal Barrier Precautions; Product Codes: ASK-45703-LDP1, ASK-45703-PBHS, ASK-45703-PHMC
Z-1995-2018
Product summary
- Event
- Event 79890
- Status
- Terminated
- Classification
- Class II
- Quantity
- 416,055 total products
- Official record key
device-enforcement:Z-1995-2018
Official wording
Reason: Product sterility may be compromised due to unsealed packaging.
Code information: Lot Numbers: Product Codes ASK-45703-LDP1: 13F17G0033 Product code ASK-45703-PBHS: 13F16M0260 13F17C0150 13F17E0406 13F17G0421 13F17H0420 13F17L0522 Product code ASK-45703-PHMC: 13F17A0047 13F17B0011 13F17H0012 13F17L0421
Distribution pattern: US Nationwide, AUSTRALIA, AUSTRIA, BELGIUM, CANADA, CZECH REPUBLIC, DENMARK, ECUADOR, FRANCE, GERMANY, GREECE, HONG KONG, INDONESIA, IRELAND, ITALY, JAPAN, KOREA, NETHERLANDS, NEW ZEALAND, PANAMA, PHILIPPINES, POLAND, PUERTO RICO, ROMANIA, SAUDI ARABIA, SINGAPORE, SLOVENIA, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, THAILAND, TRINIDAD AND TOBAGO, UNITED KINGDOM
Derived failure modes
-
Labeling or packaging
unsealed packaging
-
Sterility assurance
sterility may be compromised