AutoMate 1200 Catalog Number ODL25120; AutoMate 1250 Catalog Number ODL25125; AutoMate 2500 Catalog Number ODL25250; AutoMate 2550 Catalog Number ODL25255

Z-1169-2018

January 19, 2018

Class II

Product summary

Firm
Beckman Coulter Inc.
Event
Event 79006
Status
Completed
Classification
Class II
Quantity
98 units in the US
Official record key
device-enforcement:Z-1169-2018

Official wording

Reason: Belts on the x-axis (and on the y-axis, to a lesser extent) manufactured from 2015 are showing premature failure, posing the remote risk of cross-contamination.

Code information: All systems up to the following Serial Numbers: ODL25120 Revision T; SN3101, ODL25125 Revision W; SN1508, ODL25250 Revision T; SN2416, ODL25255 Revision W; SN4617

Distribution pattern: Australia, Austria, Belarus, Belgium, Bosnia and Herzegovina, Brazil, Canada, China, Croatia, Cyprus, Czech Republic, Denmark, Finland, France, Germany, Gibraltar, Greece, Hong Kong, Hungary, India, Ireland, Israel, Italy, Korea, Kuwait, Latvia, Lithuania, Macao, Mayotte, Mexico, Netherlands, New Zealand, Norway, Panama, Poland, Portugal, Qatar, Russia, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Turkey, United Kingdom, United States

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0