Device product

VITROS Chemistry Products Cl- Slides, Product Code 844 5207 Product Usage: For in vitro diagnostic use only. VITROS Chemistry Products Cl- Slides quantitatively measure chloride (Cl-) concentration in serum and plasma using VITROS 250/350/5,1 FS and 4600 Chemistry Systems and the VITROS 5600 Integrated System.

Z-3043-2018

July 11, 2018

Class III

Product summary

Firm
Ortho-Clinical Diagnostics
Event
Event 80697
Status
Terminated
Classification
Class III
Quantity
128196
Official record key
device-enforcement:Z-3043-2018

Official wording

Reason: The slides do not meet the maximum allowable interference (MAI) claims for triglyceride interference when using serum samples. Triglyceride interference on the Cl- Slide was observed around a triglyceride concentration 600mg/dL, however the current triglyceride interference claim states interference should not occur below 800mg/dL of triglyceride.

Code information: GEN 5 and below

Distribution pattern: Worldwide Distribution - US Nationwide and in the countries of: Australia, Belgium, Bermuda, Brazil, Canada, Chile, China, Columbia, Denmark, France, Germany, India, Italy, Japan, Mexico, Norway, Poland, Portugal, Russia, Singapore, Spain, Sweden, The Netherlands, and United Kingdom.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    The slides do not meet the maximum allowable interference (MAI) claims for triglyceride interference when using serum samples. Triglyceride interference on the Cl- Slide was observed around a triglyceride concentration 600mg/dL, however the current triglyceride interference claim states interference should not occur below 800mg/dL of triglyceride.