Nitinol Staple, Elastic Staple Large-Asymmetrical, Sterile R, Model #: a) ES-18x15x17; b) ES-18x17x19

Z-2262-2018

April 13, 2018

Class II

Product summary

Firm
TriMed Inc.
Event
Event 80099
Status
Terminated
Classification
Class II
Quantity
532
Official record key
device-enforcement:Z-2262-2018

Official wording

Reason: During a retrospective review of sterilization records, TriMed has determined that evidence of quarterly dose audits/verification required to comply with the regulations could not be found. Sterilization records indicate that all lots of staples were irradiated to at least the defined dose. There is a risk that this issue could result in patient infection if not properly sterilized.

Code information: Lot #: a) R901, 03UU, 050522, 10505-22, 0344, 0051, 11805, 00AJ, 10505-23, 050523, 180502, 180901, 180503, 18902; b) 03UV, 050523, 10505-23, 00AJ, 0052, 180901, 11809, 180902, 26408, 264-08, 11805, 180502, 118901

Distribution pattern: U.S.: IL, UT, PA, NJ, WA, TX, NV, CA, NC, MS, KS, MO, MN, FL; OUS: Australia and Sweden

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    During a retrospective review of sterilization records, TriMed has determined that evidence of quarterly dose audits/verification required to comply with the regulations could not be found. Sterilization records indicate that all lots of staples were irradiated to at least the defined dose. There is a risk that this issue could result in patient infection if not properly sterilized.