Device product

Nitinol Staple, Elastic Staple Large-Symmetrical, Sterile R, Model #: a) ES-18x14; b) ES-20x20; c) ES-25x22

Z-2263-2018

April 13, 2018

Class II

Product summary

Firm
TriMed Inc.
Event
Event 80099
Status
Terminated
Classification
Class II
Quantity
323
Official record key
device-enforcement:Z-2263-2018

Official wording

Reason: During a retrospective review of sterilization records, TriMed has determined that evidence of quarterly dose audits/verification required to comply with the regulations could not be found. Sterilization records indicate that all lots of staples were irradiated to at least the defined dose. There is a risk that this issue could result in patient infection if not properly sterilized.

Code information: Lot #: a) R900, 009J, 050511, 00AG, 0095, 183601, 10505-11, 193601; b) R903, 26404, 264-04, 03NK, 050512, 10505-12, 24604, 11811, 181102, 181103, 188803; c) 26405, 009N, 00AL, R904, 00A1, 00PN, 050513, 10505-13, 183501, 83501, 264-05, 004L,

Distribution pattern: U.S.: IL, UT, PA, NJ, WA, TX, NV, CA, NC, MS, KS, MO, MN, FL; OUS: Australia and Sweden

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    During a retrospective review of sterilization records, TriMed has determined that evidence of quarterly dose audits/verification required to comply with the regulations could not be found. Sterilization records indicate that all lots of staples were irradiated to at least the defined dose. There is a risk that this issue could result in patient infection if not properly sterilized.