Uterine Electromyographic Monitor - Monica Novii Wireless Patch System Product Usage: The Novii Interface is an accessory to the Novii Pod which provides a means of interfacing the wireless output of the Novii Pod to the transducer inputs of a CTG Fetal monitor via three cables. The Novii Interface enables signals collected by the Novii Pod to be printed and displayed on a CTG Fetal Monitor and sent on to a central network, if connected.

Z-1292-2018

December 15, 2017

Class II

Product summary

Firm
GE Healthcare, LLC
Event
Event 79464
Status
Terminated
Classification
Class II
Quantity
1742
Official record key
device-enforcement:Z-1292-2018

Official wording

Reason: if the cables of the Monica Novii System are removed by a user, and then incorrectly re-connected the Fetal Heart Rate (FHR) and Maternal Heart Rate (MHR) signals would be displayed incorrectly (swapped).

Code information: Model # - 107-PT-020 ( all units distributed from January 2016- November 2017)

Distribution pattern: Worldwide - US Nationwide Distribution.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    if the cables of the Monica Novii System are removed by a user, and then incorrectly re-connected the Fetal Heart Rate (FHR) and Maternal Heart Rate (MHR) signals would be displayed incorrectly (swapped).