Product: Milford Smartbase Model # 9934105 Pride SmartBase Pride product codes: LSMPLCG A & LSMPLCG B

Z-1420-2018

March 07, 2018

Class II

Product summary

Firm
Autochair
Event
Event 79627
Status
Ongoing
Classification
Class II
Quantity
Unavailable
Official record key
device-enforcement:Z-1420-2018

Official wording

Reason: Recalling firm was informed that as a result of combination of a weak spring and incorrect alignment of the locking pin for latching the leg positions, product legs folded inwards while lifting a person off a bed and the individual was dropped.

Code information: All Lots

Distribution pattern: PA, and worldwide

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Recalling firm was informed that as a result of combination of a weak spring and incorrect alignment of the locking pin for latching the leg positions, product legs folded inwards while lifting a person off a bed and the individual was dropped.