Device product
DigitalDiagnost Release 3.1 (Stitching Patient Support) 712026
Z-2862-2018
Product summary
- Event
- Event 80577
- Status
- Terminated
- Classification
- Class II
- Quantity
- Unavailable
- Official record key
device-enforcement:Z-2862-2018
Official wording
Reason: The hook does not securely hold the footplate in the vertical position. Therefore, the footplate can fall or swing down and injury operator or patient.
Code information: Serial Numbers: SN14000256 SN14000250 SN15000024 SN14000170 SN14000052 SN14000224 SN15000039 SN14000144 SN14000253 SN14000239 SN14000186 SN14000075 SN14000193 SN14000254 SN15000001 SN14000089 SN14000223 SN14000267 SN14000238 SN14000164 SN14000165 SN14000200 SN14000005 SN14000040 SN15000015 SN15000029 SN14000120
Distribution pattern: US Nationwide; International to 66 countries
Derived failure modes
-
Unknown
The hook does not securely hold the footplate in the vertical position. Therefore, the footplate can fall or swing down and injury operator or patient.