Device product
Trident Endoscopic Ultrasonic Aspiration Needle (FNA)
Z-1878-2018
Product summary
- Event
- Event 79911
- Status
- Terminated
- Classification
- Class II
- Quantity
- 40 total
- Official record key
device-enforcement:Z-1878-2018
Official wording
Reason: The combination of the firmness of the mass, the potential endoscope and elevator position combined with the force required to penetrate the mass may contribute to the breakage of the needle in the human body, which could cause injury.
Code information: Identifier EUS-22-0-N UPN ES34601 M171226172
Distribution pattern: US distribution to one distributor in MI.
Derived failure modes
-
Unknown