Skip to content
Recall Observatory FDA recall evidence

Device product

Ceterix Disposable Knot Pusher / Suture Cutter; NovoCut Suture Manager. Model# CTX-C001. orthopedic arthroscopic accessory.

Z-2945-2018

May 16, 2018

Class II

Product summary

Firm
Ceterix Orthopedics, Inc.
Event
Event 80688
Status
Terminated
Classification
Class II
Quantity
96 units
Official record key
device-enforcement:Z-2945-2018

Official wording

Reason: Possible nonconforming blades may result in a higher than specified cut force required to cut the suture.

Code information: Lot#M180018

Distribution pattern: United States distribution.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Possible nonconforming blades may result in a higher than specified cut force required to cut the suture.