Skip to content
Recall Observatory FDA recall evidence

Device product

Uterine Electromyographic Monitor -Monica IF24 Interface System Product Usage: The Monica IF24 CTG Interface Device is an accessory to the Monica AN24 which provides a means of interfacing the wireless output of the Monica AN24 to the transducer inputs of a CTG Fetal monitor. The Monica IF24 CTG Interface Device enables signals collected by the Monica AN24 to be printed and displayed on a CTG Fetal Monitor and sent on to a central network, if connected. The device is intended for use by healthcare professionals in a clinical setting.

Z-1293-2018

December 15, 2017

Class II

Product summary

Firm
GE Healthcare, LLC
Event
Event 79464
Status
Terminated
Classification
Class II
Quantity
355
Official record key
device-enforcement:Z-1293-2018

Official wording

Reason: if the cables of the Monica IF24 System are removed by a user, and then incorrectly re-connected the Fetal Heart Rate (FHR) and Maternal Heart Rate (MHR) signals would be displayed incorrectly (swapped).

Code information: Model # - 105-PT-001 ( all units distributed from October 2011 - May 2016)

Distribution pattern: Worldwide - US Nationwide Distribution.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    if the cables of the Monica IF24 System are removed by a user, and then incorrectly re-connected the Fetal Heart Rate (FHR) and Maternal Heart Rate (MHR) signals would be displayed incorrectly (swapped).