J-Plasma Precise 360(R), Catalog Numbers: BVX-330BR, BVX-330NR, BVX-450BR, BVX-450NR
Z-1156-2018
Product summary
- Event
- Event 79398
- Status
- Terminated
- Classification
- Class II
- Quantity
- 595 units
- Official record key
device-enforcement:Z-1156-2018
Official wording
Reason: Bovie Medical is issuing a correction to address incomplete insertion of the J-Plasma(R) handpiece cable plug into the generator receptacle resulting in an incomplete helium seal. The potential exists for patient fluid backflow into the handpiece and the generator receptacle leading to risk of cross-contamination if the generator is reused.
Code information: All lots manufactured since 01/01/2014
Distribution pattern: worldwide
Derived failure modes
-
Unknown
Bovie Medical is issuing a correction to address incomplete insertion of the J-Plasma(R) handpiece cable plug into the generator receptacle resulting in an incomplete helium seal. The potential exists for patient fluid backflow into the handpiece and the generator receptacle leading to risk of cross-contamination if the generator is reused.