Device product
Optima CT 520
Z-2585-2018
Product summary
- Event
- Event 80484
- Status
- Terminated
- Classification
- Class II
- Quantity
- 42 units total
- Official record key
device-enforcement:Z-2585-2018
Official wording
Reason: Some CT and CT/PET systems recently installed with a GE supplied A! electrical panel may not be properly wired to the partial system UPS (Uninteruptible Power Supply).
Code information: Serial number, CBCRG1700051HM, System ID Number 580622O520, UDI Number 01008406821025681117120021CBCRG1700051HM
Distribution pattern: Nationwide distribution to AL, CA, CO, FL, IL, IN, MA, MD, MI, MS, NC, NJ, NY,OH, OK, SC, TN, TX, VA.
Derived failure modes
-
Unknown
Some CT and CT/PET systems recently installed with a GE supplied A! electrical panel may not be properly wired to the partial system UPS (Uninteruptible Power Supply).