Device product
Insulia Diabetes Management Companion
Z-0478-2019
Product summary
- Firm
- Voluntis
- Event
- Event 81508
- Status
- Terminated
- Classification
- Class II
- Quantity
- 14
- Official record key
device-enforcement:Z-0478-2019
Official wording
Reason: The firm identified an issue with the basal calculator identified on the Android version of the Insulia application. The bug could lead from low-impact to high-impact hypoglycemia depending on the circumstances. A new update of Insulia is now available (version 1.6.42), which resolves this issue.
Code information: Software version numbers 1.4.3 / 1.6.3 / 1.6.31 / 1.6.40
Distribution pattern: Distributed to accounts in Arizona, Masachusetts, Michigan, Ohio, Tennessee, and Vermont. International distribution to Canada and France (however, French customers are not affected by this recall).
Derived failure modes
-
Unknown
The firm identified an issue with the basal calculator identified on the Android version of the Insulia application. The bug could lead from low-impact to high-impact hypoglycemia depending on the circumstances. A new update of Insulia is now available (version 1.6.42), which resolves this issue.