Duo-Flow Double Lumen Soft Tip CRRT/ABP Catheter Set; DL11/24 . Hemodialysis catheter insertion kits
Z-2934-2018
Product summary
- Event
- Event 80581
- Status
- Terminated
- Classification
- Class II
- Quantity
- N/A
- Official record key
device-enforcement:Z-2934-2018
Official wording
Reason: The suture wings have the potential to crack or break. Potential patient injury includes catheter dislocation, bleeding, and air emboli. The firm has received six complaints (29 broken suture wings) to date for malfunction, without serious injury or death.
Code information: MCAL840 MCBB800 MCBW450 MCCH690
Distribution pattern: Nationwide distribution; worldwide distribution.
Derived failure modes
-
Unknown
The suture wings have the potential to crack or break. Potential patient injury includes catheter dislocation, bleeding, and air emboli. The firm has received six complaints (29 broken suture wings) to date for malfunction, without serious injury or death.