Adult AirLife dual-limb, dual-heat circuit (Model AH280) with AirLife Humidification Chamber, Model Number AH290. The AH290 Humidification Chamber and Heated Breathing Circuit Kit(s) operate as part of a system for delivering health and humidified respirator gasses to patients requiring invasive and non-invasive ventilation.

Z-0346-2018

November 29, 2014

Class I

Product summary

Firm
Vyaire Medical
Event
Event 78636
Status
Terminated
Classification
Class I
Quantity
16,670 units in total
Official record key
device-enforcement:Z-0346-2018

Official wording

Reason: Manufacturing error in humidification chamber may cause an overfill/overflow condition of the chamber. The may case a water back-up into the patient circuit resulting in an excessive amount of water entering the airway/lungs of the ventilated patient.

Code information: 0001100127 0001116871 0001116872 0001124936 0001129065 0001129066 0001129067 0001129069 0001129070 0001131426 0001131428 0001134350 0001135400 0001136027 0001136028 0001136029 0001145151 0001145152 0001145153 0001157125

Distribution pattern: US Nationwide Distribution

Derived failure modes

  • Manufacturing or process control

    reason.manufacturing_process_control · v1.0.0

    Manufacturing error