Skip to content
Recall Observatory FDA recall evidence

Device product

Vereos PET/CT, Software version 2.0.1. Model Number: 882446 Product Usage: The Philips Vereos PET/CT system is a diagnostic imaging device that combines Positron Emission Tomography (PET) and X-ray Computed Tomography (CT) systems. The CT subsystem images anatomical cross-sections by computer reconstruction of X-ray transmission data. The PET subsystem images the distribution of PET anatomy-specific radiopharmaceuticals in the patient. The PET/CT system is used for the purposes of detecting, localizing, diagnosing, staging, re-staging, and follow-up for monitoring therapy response of various diseases in oncology, cardiology, and neurology. The system is intended to image the whole body, heart, brain, lung, gastrointestinal, bone, lymphatic, and other major organs for a wide range of patient types, sizes, and extent of diseases. Both subsystems can also be operated as fully functional, independent diagnostic tools, including application of the CT scanner for diagnosis and for use in radiation therapy planning.

Z-0302-2019

September 04, 2018

Class II

Product summary

Firm
Philips Medical Systems (Cleveland) Inc
Event
Event 81173
Status
Terminated
Classification
Class II
Quantity
61 units
Official record key
device-enforcement:Z-0302-2019

Official wording

Reason: Software Correction: Software errors that may result in incorrect 4D CT images, PET images with SUV quantification error, or disruption of interventional CT procedures; and Software errors that may result in partial set of images, inability to generate CT or PET images, incorrect scan parameters, or scanning the incorrect portion of the body.

Code information: Serial Numbers: 900004 900005 900006 900007 900008 900009 900010 900011 900012 900013 900014 900015 900016 900017 900018 900019 900020 900021 900022 900023 900024 900025 900026 900027 900028 900029 900030 900031 900032 900033 900034 900035 900036 900037 900038 900039 900040 900041 900042 900043 900044 900045 900046 900047 900048 900049 900050 900051 900052 900053 900054 900055 900056 900058 900059 900060 900062 900064 900065 900066 900067

Distribution pattern: US Nationwide Distribution

Derived failure modes

  • Device software or design

    reason.device_software_design · v1.0.0

    Software Correction: Software errors that may result in incorrect