Device product

QUADROX-i Neonatal without integrated arterial filter and VHK 11000 (VKMO 10000-USA#QUADROX-i Neo.+Reservoir), Part Number 701055120 The membrane oxygenator QUADROXi Neonatal is intended for use in extracorporeal circulation during cardiopulmonary bypass in cardiac surgery.

Z-0436-2019

October 25, 2018

Class I

Product summary

Firm
Maquet Cardiovascular Us Sales, Llc
Event
Event 81382
Status
Terminated
Classification
Class I
Quantity
490 total in US
Official record key
device-enforcement:Z-0436-2019

Official wording

Reason: There is a potential for blood leakage at the blood inlet connector when using the reducing screw adapter 1/4 x 3/16 that may result in clinically relevant loss of prime before patient connection or blood loss during patient connection.

Code information: UDI: 04037691780139; Lots: 92212343, 92214685, 92225187

Distribution pattern: Worldwide Distribution: US (nationwide) to states of: AZ, CA, FL, IA, LA, NC, and TX and countries of: Canada, Argentina, Brazil, Chile, Colombia, Mexico, Uruguay, United Arab Emirates, Belgium, Switzerland, Germany, Egypt, Spain, France, United Kingdom, Ireland, Italy, Lebanon, Netherlands, Pakistan, Portugal, Reunion, Saudi Arabia, Sweden, South Africa, Poland, Serbia, Russia, Turkey, Ukraine, China, India, Japan, Republic of Korea (South Korea), Malaysia, Thailand, and Taiwan.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    There is a potential for blood leakage at the blood inlet connector when using the reducing screw adapter 1/4 x 3/16 that may result in clinically relevant loss of prime before patient connection or blood loss during patient connection.