Device product
VNS Therapy Programmer, Model 3000, v1.0 System
Z-2255-2018
Product summary
- Firm
- Cyberonics, Inc
- Event
- Event 80062
- Status
- Terminated
- Classification
- Class II
- Quantity
- 252
- Official record key
device-enforcement:Z-2255-2018
Official wording
Reason: Certain Model 3000 programming events can result in miscalculation of parameters stored in the Models 103, 104, 105, and 106 generators. During these programming events, the miscalculations can lead to: " Delivery of more stimulation than intended, resulting in painful stimulation or other common side effects (Model 106 only); " No stimulation in the case of device disablement (Burst Watchdog Timeout), resulting in no therapy to the patient (Model 106 only); " Delivery of less stimulation than intended, resulting in therapeutic settings not being achieved within device specification (Models 103, 104, 105, or 106); and/or " Delays or absence of the 75% and 50% battery life indicators displayed by the programming software (Models 103, 104, 105, or 106).
Code information: Serial Numbers: FMT77F2, FKR57F2, FCR77F2, CJR77F2, BNT77F2, BNS77F2, 7XR77F2, 7XP57F2, 79R77F2, 90S77F2, 97Q57F2, 6NT77F2, 9GP57F2, 6KQ57F2, B7Q57F2, 57P57F2, 55T77F2, J6R57F2, HWR77F2, 1G7G1F2, 1JR57F2, 9XP57F2, 4LR77F2, 4GT77F2, 4GR57F2, 2DT77F2, G2S77F2, FVQ77F2, G5T77F2, FDQ57F2, F5R57F2, F2R57F2, F1Q57F2, GHS77F2, DVR77F2, DNP57F2, DMYB7F2, DJS77F2, D1R57F2, CVQ57F2, H4R57F2, 85S77F2, 89T77F2, 8LYB7F2, 97YB7F2, 6MQ57F2, 6CQ57F2, 5HS77F2, JLQ57F2, JGS77F2, 11R57F2, 1TQ57F2, 1YR77F2, 2DR77F2, ,FXP57F2 , FTR77F2, FPQ57F2, FNQ57F2, FLYB7F2, FFP57F2, GGT77F2, GSQ57F2, CMS77F2, H1R77F2, H3R57F2, 7LS77F2, 9KR57F2, 9NS77F2, 65R57F2, B8R77F2, B9T77F2, 5D0C7F2, B9YB7F2, BJR57F2, JCQ57F2, J9YB7F2, HMYB7F2, CDQ57F2, FXQ57F2, G5S77F2, G3R77F2, FRQ57F2, G7YB7F2, FLS77F2, G90C7F2, FGR77F2, FFT77F2, G9Q57F2, FCT77F2, G9R57F2, F9YB7F2, F9S77F2, F8T77F2, GDQ57F2, F7R57F2, F4T77F2, DWR77F2, DTR77F2, DSQ57F2, GMQ57F2, DLS77F2, DH0C7F2, DGS77F2, DGP57F2, GZR77F2, D1Q57F2, CSQ57F2, CLS77F2, C9YB7F2, C70C7F2, H1R57F2, C1R77F2, 7MS77F2, 78YB7F2, 9SQ57F2, 5VQ57F2, 5RQ57F2, B60C7F2, 5NQ57F2, 5MR77F2, 5KS77F2, B8T77F2, BHP57F2, BJR77F2, 4JS77F2, H4T77F2, JZQ57F2, JMS77F2, JMQ57F2, JG0C7F2, JBQ57F2, J8T77F2, ,HZQ57F2, J7R57F2, HVR77F2, HTP57F2, HNS77F2, HMQ57F2, HKQ57F2, H7S77F2, D3R57F2, 1CQ57F2, 1HT77F2, 1MYB7F2, 2CP57F2, 2LYB7F2, CLYB7F2, CMYB7F2, B4R77F2, B2S77F2, B0R57F2, 9QS77F2, D8Q57F2, 960C7F2, 95T77F2, 8JR77F2, 8HR57F2, BXP57F2, BWR77F2, BTQ77F2, C8YB7F2, C60C7F2, BLR77F2, BLQ57F2, BGR77F2, BGP57F2, CHQ57F2, CKR57F2, CPP57F2, CSR77F2, CTQ77F2, CXR77F2, 9MS77F2, 9JR77F2, 9DQ57F2, D7S77F2, 94T77F2, 8PQ57F2, BWQ77F2, BSR77F2, BNQ57F2, C9R57F2, C8R57F2, C5R57F2, CF0C7F2, C3R57F2, CFP57F2, BDT77F2, B9S77F2, B5R57F2, CPQ57F2, B4R57F2, D2R57F2, 9KQ57F2, D5R77F2, 9F0C7F2, 980C7F2, D9YB7F2, 8WR77F2, DB0C7F2, 9P7G1F2, D3S77F2, D6R57F2, D7R57F2, 9CR77F2, 9G0C7F2, 96R57F2, DFP57F2, C0S77F2, BN7G1F2, BQ7G1F2, C8S77F2, CB0C7F2, CCR77F2, BHR57F2, CHP57F2, BCR77F2, BBR77F2, BB0C7F2, B90C7F2, B70C7F2, B2R57F2, 9VP57F2, 9TQ57F2, 9GR57F2, 9FP57F2, 9BS77F2, ,98R57F2 ,D80C7F2, 970C7F2, D8YB7F2, 95R77F2, 8YR77F2, C1R57F2, C80C7F2, C3R77F2, CNS77F2, 9VQ77F2, D4T77F2, 9DR77F2, 91R57F2, 8VQ77F2, 8RQ57F2, DBS77F2, 8JS77F2, C0R57F2, BTR77F2, BMYB7F2, BKR77F2, BBP57F2 and B8S77F2
Distribution pattern: AL, CA, CO, DE, FL, GA, ID, IL, IN, ME, MO, MS, NC, NJ, NY, PA, TN, TN, TX, UT, WA and WI
Derived failure modes
-
Unknown
Certain Model 3000 programming events can result in miscalculation of parameters stored in the Models 103, 104, 105, and 106 generators. During these programming events, the miscalculations can lead to: " Delivery of more stimulation than intended, resulting in painful stimulation or other common side effects (Model 106 only); " No stimulation in the case of device disablement (Burst Watchdog Timeout), resulting in no therapy to the patient (Model 106 only); " Delivery of less stimulation than intended, resulting in therapeutic settings not being achieved within device specification (Models 103, 104, 105, or 106); and/or " Delays or absence of the 75% and 50% battery life indicators displayed by the programming software (Models 103, 104, 105, or 106).