Device product
CT Goldseal Optima CT 600
Z-2583-2018
Product summary
- Event
- Event 80484
- Status
- Terminated
- Classification
- Class II
- Quantity
- 42 units total
- Official record key
device-enforcement:Z-2583-2018
Official wording
Reason: Some CT and CT/PET systems recently installed with a GE supplied A! electrical panel may not be properly wired to the partial system UPS (Uninteruptible Power Supply).
Code information: Serial Number 51941YC3, System ID Number 406434660
Distribution pattern: Nationwide distribution to AL, CA, CO, FL, IL, IN, MA, MD, MI, MS, NC, NJ, NY,OH, OK, SC, TN, TX, VA.
Derived failure modes
-
Unknown
Some CT and CT/PET systems recently installed with a GE supplied A! electrical panel may not be properly wired to the partial system UPS (Uninteruptible Power Supply).