Device product
Edwards Lifesciences Thin-Flex Single Stage Venous Drainage Cannula, 34F (11.3MM) X 40CM (16"), resterilized.
Z-3030-2018
Product summary
- Event
- Event 80317
- Status
- Terminated
- Classification
- Class II
- Quantity
- 5 units total
- Official record key
device-enforcement:Z-3030-2018
Official wording
Reason: The expiration date on the device labeling exceeds its actual validated shelf life because the sterilization process used does not have adequate support for the extended shelf life.
Code information: Code Number TF034L-R; Lot Number 59739865; Serial Numbers 18304192017-12, 18304192017-13, 18304192017-14, 18304192017-15, 18304192017-16
Distribution pattern: US distribution to AZ, CA, FL, OH, PA, TX, VA, WV
Derived failure modes
-
Unknown
The expiration date on the device labeling exceeds its actual validated shelf life because the sterilization process used does not have adequate support for the extended shelf life.