Device product
Duo-Flow XTP Straight Full Set; Hemodialysis catheter insertion kits, AAC02106 AAC02107 AAC02108
Z-2932-2018
Product summary
- Event
- Event 80581
- Status
- Terminated
- Classification
- Class II
- Quantity
- N/A
- Official record key
device-enforcement:Z-2932-2018
Official wording
Reason: The suture wings have the potential to crack or break. Potential patient injury includes catheter dislocation, bleeding, and air emboli. The firm has received six complaints (29 broken suture wings) to date for malfunction, without serious injury or death.
Code information: MCAW430, MCBR560, MCCX620 MCCF020 MCFK440 MKAY620
Distribution pattern: Nationwide distribution; worldwide distribution.
Derived failure modes
-
Unknown
The suture wings have the potential to crack or break. Potential patient injury includes catheter dislocation, bleeding, and air emboli. The firm has received six complaints (29 broken suture wings) to date for malfunction, without serious injury or death.