Device product
Hamilton-C3 Ventilator Unit Product Usage: The HAMILTON-C3 ventilator is intended to provide positive pressure ventilator support to adults, pediatrics, infants and neonates depending on its configuration. The device is used in the intensive care ward, intermediate care ward, emergency ward, long term acute care hospital or in the recovery room, and during transfer of ventilated patients within the hospital.
Z-2796-2018
Product summary
- Event
- Event 80467
- Status
- Terminated
- Classification
- Class II
- Quantity
- 2 devices
- Official record key
device-enforcement:Z-2796-2018
Official wording
Reason: An out of specification component on the control board may cause the ventilator to open the inspiratory channel and the expiratory valve. The ventilator will shut down and an emergency buzzer providing an acoustic alarm and the blinking alarm LED on the device s front side.
Code information: Serial Number: 3084, 3088
Distribution pattern: US distribution to CA, CO, IL, LA, MN, NY, OK, VA, WI. Worldwide distribution to China, Germany, India, Indonesia, Israel, Kazakhstan, Korea, Pakistan, Taiwan, Poland, Armenia.
Derived failure modes
-
Potency or specification failure
out of specification