Device product
Haemonetics¿ OrthoPAT¿ System-OPAT QuickConnect Reservoirs. Product Code: OPT-R-1000 The OrthoPAT orthopedic perioperative autotransfusion system indicated for use to salvage RBCs from blood lost intraoperatively and postoperatively during surgical procedures
Z-2951-2018
Product summary
- Event
- Event 80740
- Status
- Terminated
- Classification
- Class II
- Quantity
- 348 units
- Official record key
device-enforcement:Z-2951-2018
Official wording
Reason: OrthoPAT¿ reservoirs are missing the cannula inside the reservoirs. The absence of this cannula prevents the emptying of blood from the reservoirs. The inability to empty the reservoir means that the collected blood cannot be processed.
Code information: Lot # 1117066
Distribution pattern: US Nationwide Distribution
Derived failure modes
-