Device product
e600 Automatic Transport Ventilator, Model Number 01EVE600
Z-0977-2019
Product summary
- Event
- Event 82178
- Status
- Terminated
- Classification
- Class I
- Quantity
- 119
- Official record key
device-enforcement:Z-0977-2019
Official wording
Reason: There is a potential for the unit to catch fire due to one of the screen retaining screws being incorrectly inserted. Due to this improper screw insertion the screw sat higher than normal allowing it to contact the underside of the main board causing the screw to rub against the bottom side of the board, wearing away the insulation layer and creating a short circuit.
Code information: EV60001 to EV60126
Distribution pattern: The products were distributed to the following US states: AZ, CA, CO, FL, OH, PA, TX, and VA.
Derived failure modes
-
Unknown
There is a potential for the unit to catch fire due to one of the screen retaining screws being incorrectly inserted. Due to this improper screw insertion the screw sat higher than normal allowing it to contact the underside of the main board causing the screw to rub against the bottom side of the board, wearing away the insulation layer and creating a short circuit.