SPECTRUM IQ INFUSION SYSTEM WITH DOSE IQ SAFETY SOFTWARE, 3570009
Z-0107-2019
Product summary
- Event
- Event 81158
- Status
- Terminated
- Classification
- Class II
- Quantity
- 354 units
- Official record key
device-enforcement:Z-0107-2019
Official wording
Reason: If an operator attempts to manually program a Basic Infusion for a drug that is not included in the Drug Library using the Basic-Dose Mode, the operator will receive a false hard limit exceeded error, preventing him or her from programming the infusion.
Code information: UDI 00085412610900 All Serial Numbers distributed prior to 07/09/2018
Distribution pattern: NJ
Derived failure modes
-
Unknown
If an operator attempts to manually program a Basic Infusion for a drug that is not included in the Drug Library using the Basic-Dose Mode, the operator will receive a false hard limit exceeded error, preventing him or her from programming the infusion.