Device product
Cardiosave Rescue IABP, Part Number 0998-UC-0800-83
Z-0100-2019
Product summary
- Event
- Event 81055
- Status
- Ongoing
- Classification
- Class I
- Quantity
- N/A
- Official record key
device-enforcement:Z-0100-2019
Official wording
Reason: There is a potential for interruption and/or inability to start therapy to the patient prior to or during use of Cardiosave IABP for users who are at altitudes above 3200 Ft.
Code information: UDI - 10607567108407 All Serial Numbers
Distribution pattern: US Nationwide; Internationally to 77 countries
Derived failure modes
-
Unknown
There is a potential for interruption and/or inability to start therapy to the patient prior to or during use of Cardiosave IABP for users who are at altitudes above 3200 Ft.