Spine & Trauma 3D Navigation 1.0 (subpart of the system Navigation Software Spine & Trauma 3D, Version 3.0 ) Intended Use The Spine & Trauma Navigation System is intended as an intraoperative image-guided localization system to enable minimally invasive surgery.

Z-1082-2019

February 27, 2019

Class I

Product summary

Firm
Brainlab AG
Event
Event 82304
Status
Terminated
Classification
Class I
Quantity
148 units
Official record key
device-enforcement:Z-1082-2019

Official wording

Reason: In certain occurrences, the affected navigation software application might unexpectedly display a navigated instrument in an axial, coronal/sagittal (ACS) view representation with fixed planes in the image reconstruction (the not-updated ACS view ), instead of displaying the desirable view representation Inline View , which is commonly used for navigating invasive instruments at the spine. This could lead the surgeon to be unable to determine the position of the navigated instrument. This might occur after a crash restore or after changing between different navigation workflows during the same patient treatment.

Code information: Version 1.0.0.249, GTIN 04056481140328

Distribution pattern: Worldwide Distribution: US (Nationwide) and countries of: Australia, Austria, Belgium, Canada, Czech Republic, Denmark, Ecuador, France, Germany, Greece, Hong Kong, Ireland, Italy, Japan, Kazakhstan, Malaysia, New Zealand, Philippines, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand and United Kingdom.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    In certain occurrences, the affected navigation software application might unexpectedly display a navigated instrument in an axial, coronal/sagittal (ACS) view representation with fixed planes in the image reconstruction (the not-updated ACS view ), instead of displaying the desirable view representation Inline View , which is commonly used for navigating invasive instruments at the spine. This could lead the surgeon to be unable to determine the position of the navigated instrument. This might occur after a crash restore or after changing between different navigation workflows during the same patient treatment.