Device product

STEPHANIX D2RS Digital Dynamic Remote Systemhttp://cts.fda.gov/division-tracking/images/trash.png Product The D2RS Digital Dynamic Remote System is indicated for use in generating fluoroscopic images of human anatomy for vascular angiography, diagnostic and interventional procedures. It is also indicated for generating fluoroscopic images of human anatomy for cardiology, diagnostic, and interventional procedures. It is intended to replace fluoroscopic images obtained through image intensifier technology. Not intended for mammography applications.

Z-0557-2019

November 02, 2018

Class II

Product summary

Firm
Stephanix
Event
Event 81582
Status
Terminated
Classification
Class II
Quantity
17 units
Official record key
device-enforcement:Z-0557-2019

Official wording

Reason: The firm has detected a potential risk using the command. After releasing the command, the movement of the table may continue instead of stopping.

Code information: DHJ008, DIG010, DIH012, DIH014, DIK022, DIK024, DIL025, DIL030, DJK051, DJK052, DKI032, DKI033, DKJ047, DKK051, DKK052, DLA001, DML062