Device product
Discovery NM/CT 670 DR, model 5376204-70-56 intended for use in detecting, localizing, diagnosing of diseases and organ function for the evaluation of diseases, trauma, abnormalities, and disorders such as, but not limited to, cardiovascular disease, neurological disorders and cancer.
Z-0606-2019
Product summary
- Event
- Event 81680
- Status
- Completed
- Classification
- Class II
- Quantity
- 4 units
- Official record key
device-enforcement:Z-0606-2019
Official wording
Reason: When your system was installed, the Main disconnect switch, EPO button on the A1 panel, and Remote EPO button(s) (outside of the A1 panel) may not have been wired correctly which could result in these not shutting off all power to the entire system as intended, resulting in a potential electric hazard.
Code information: GTIN 00840682124393 Serial Numbers: DDRW34032, DDRX34038, DDRX34041, DDRX34045
Distribution pattern: CA, DC, OH, UT, WI, Canada
Derived failure modes
-
Unknown
When your system was installed, the Main disconnect switch, EPO button on the A1 panel, and Remote EPO button(s) (outside of the A1 panel) may not have been wired correctly which could result in these not shutting off all power to the entire system as intended, resulting in a potential electric hazard.