Device product
QUADROX-i Neonatal with integrated arterial filter and VHK 11000 (VKMO 11000-USA#QUADROX-i Neo.+Reservoir), Part Number 701055207 The membrane oxygenator QUADROXi Neonatal is intended for use in extracorporeal circulation during cardiopulmonary bypass in cardiac surgery.
Z-0437-2019
Product summary
- Event
- Event 81382
- Status
- Terminated
- Classification
- Class I
- Quantity
- 490 total in US
- Official record key
device-enforcement:Z-0437-2019
Official wording
Reason: There is a potential for blood leakage at the blood inlet connector when using the reducing screw adapter 1/4 x 3/16 that may result in clinically relevant loss of prime before patient connection or blood loss during patient connection.
Code information: UDI: 04037691773414; Lots: 92209374, 92211752, 92215106, 92222131, 92223169, 92228805, 92234286, 92236876
Distribution pattern: Worldwide Distribution: US (nationwide) to states of: AZ, CA, FL, IA, LA, NC, and TX and countries of: Canada, Argentina, Brazil, Chile, Colombia, Mexico, Uruguay, United Arab Emirates, Belgium, Switzerland, Germany, Egypt, Spain, France, United Kingdom, Ireland, Italy, Lebanon, Netherlands, Pakistan, Portugal, Reunion, Saudi Arabia, Sweden, South Africa, Poland, Serbia, Russia, Turkey, Ukraine, China, India, Japan, Republic of Korea (South Korea), Malaysia, Thailand, and Taiwan.
Derived failure modes
-
Unknown
There is a potential for blood leakage at the blood inlet connector when using the reducing screw adapter 1/4 x 3/16 that may result in clinically relevant loss of prime before patient connection or blood loss during patient connection.