Device product
OviTex Reinforced BioScaffold 6x10cm, Part Number F10254-0610G
Z-0416-2019
Product summary
- Event
- Event 81166
- Status
- Terminated
- Classification
- Class II
- Quantity
- 881 total
- Official record key
device-enforcement:Z-0416-2019
Official wording
Reason: Degradation of the PGA suture material used in the manufacture of the resorbable mesh devices was observed during an on-going product stability study. Further investigation indicated that devices over 18-months showed evidence of no longer meeting the pre-defined ball burst specification respective to the number of tissue layers.
Code information: ERT-6H12
Distribution pattern: The products were distributed to the following US states: AL, CA, FL, IN, MA, MI, NH, and NY.
Derived failure modes
-
Unknown
Degradation of the PGA suture material used in the manufacture of the resorbable mesh devices was observed during an on-going product stability study. Further investigation indicated that devices over 18-months showed evidence of no longer meeting the pre-defined ball burst specification respective to the number of tissue layers.