Device product
Xcela Power Injectable PICC with PASV (XCELA PASV 4F SL 55CM Catheter Kit CE PG), Cat no. 55-711
Z-0595-2019
Product summary
- Event
- Event 81477
- Status
- Terminated
- Classification
- Class II
- Quantity
- Box of 1 Units
- Official record key
device-enforcement:Z-0595-2019
Official wording
Reason: Specific lots of Xcela Power Injectable and BioFlo PICCs are labelled with incorrect expiration dates. The labelled expiration date extends the shelf life of the product beyond the date supported by validation testing. Although the product under recall has not yet reached its validated expiry date, removal of the affected product is warranted to prevent use beyond the validated expiry date.
Code information: Lot 5096448
Distribution pattern: Distributed to accounts in Florida. Foreign distribution to Great Britain.
Derived failure modes
-
Unknown
Specific lots of Xcela Power Injectable and BioFlo PICCs are labelled with incorrect expiration dates. The labelled expiration date extends the shelf life of the product beyond the date supported by validation testing. Although the product under recall has not yet reached its validated expiry date, removal of the affected product is warranted to prevent use beyond the validated expiry date.