Skip to content
Recall Observatory FDA recall evidence

Device product

Siemens ADVIA Centaur aTG (100 test kit) Siemens Material Number (SMN): 10492398

Z-1233-2019

December 07, 2018

Class II

Product summary

Firm
Siemens Healthcare Diagnostics, Inc
Event
Event 82521
Status
Terminated
Classification
Class II
Quantity
10673 kits
Official record key
device-enforcement:Z-1233-2019

Official wording

Reason: Positive bias anti-thyroglobulin with ADVIA Centaur aTG kit lots ending in 316 and lower and Atellica IM aTG kit lots ending in 317 and lower when compared to the standardization to World Health Organization (WHO) Reference Preparation MRC 65/93 stated in the Instructions for Use (IFU).

Code information: Lot Number/Exp. Date: 68748302 12/13/2018 68749302 12/13/2018 83577304 1/12/2019 88637304 1/12/2019 5610306 2/11/2019 96436306 2/11/2019 19097308 3/14/2019 31866310 4/27/2019 45386310 4/27/2019 55627312 5/27/2019 69208316 6/29/2019 73000316 6/29/2019

Distribution pattern: Nationwide Foreign: Belgium¿¿¿¿¿¿¿¿ Croatia¿¿¿¿¿¿¿¿ Czech¿Republic¿ Denmark¿¿¿¿¿¿¿¿ France¿¿¿¿¿¿¿¿¿ Germany¿¿¿¿¿¿¿¿ Greece¿¿¿¿¿¿¿¿¿ Ireland¿¿¿¿¿¿¿¿ Italy¿¿¿¿¿¿¿¿¿¿ Poland¿¿¿¿¿¿¿¿¿ Portugal¿¿¿¿¿¿¿ Slovakia¿¿¿¿¿¿¿ Spain¿¿¿¿¿¿¿¿¿¿ Switzerland¿¿¿¿ Turkey¿¿¿¿¿¿¿ BOGOTA QUARRY BAY, HONG KONG YANGON SANTIAGO DE CHILE MONTEVIDEO BUENOS AIRES LIMA JOINVILLE, SC, BRAZIL Selangor Darul Ehsan Dhaka Ha Noi City Gurgaon, Haryana Hanoi-Hoangmai ASUNCION SHANGHAI Seoul 120 SINGAPORE RIYADH 11411 TAIPEI R.O.C. QUITO HUNTINGWOOD, NEW SOUTH WALES ZAPOPAN Ho Chi Minh-District 3 Huaykwang, Bangkok RIYADH SAN JOSE CHIBA LAVAL HO CHI MINH CITY MANAGUA Gurgaon, Haryana

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Positive bias anti-thyroglobulin with ADVIA Centaur aTG kit lots ending in 316 and lower and Atellica IM aTG kit lots ending in 317 and lower when compared to the standardization to World Health Organization (WHO) Reference Preparation MRC 65/93 stated in the Instructions for Use (IFU).