Device product
Discovery NM/CT 670 ES, model 5376204-70-57
Z-0607-2019
Product summary
- Event
- Event 81680
- Status
- Completed
- Classification
- Class II
- Quantity
- 2 units
- Official record key
device-enforcement:Z-0607-2019
Official wording
Reason: When your system was installed, the Main disconnect switch, EPO button on the A1 panel, and Remote EPO button(s) (outside of the A1 panel) may not have been wired correctly which could result in these not shutting off all power to the entire system as intended, resulting in a potential electric hazard.
Code information: GTIN 00840682123877 Serial Numbers: ESDX35013, ESDW35010
Distribution pattern: CA, DC, OH, UT, WI, Canada
Derived failure modes
-
Unknown
When your system was installed, the Main disconnect switch, EPO button on the A1 panel, and Remote EPO button(s) (outside of the A1 panel) may not have been wired correctly which could result in these not shutting off all power to the entire system as intended, resulting in a potential electric hazard.