Skip to content
Recall Observatory FDA recall evidence

Device product

Edwards Lifesciences Swan Ganz Thermodilution Catheter, REF 151F7 Product Usage: The Swan-Ganz Thermodilution Catheter provides a diagnostic tool for physicians to rapidly determine hemodynamic pressures and cardiac output when used with a compatible cardiac output computer. Swan-Ganz thermodilution catheters are indicated for the assessment of a patient s hemodynamic condition through direct intra-cardiac and pulmonary artery pressure monitoring, cardiac output determination, and for infusing solutions. The distal (pulmonary artery) port also allows sampling of mixed venous blood for the assessment of oxygen transport balance and the calculation of derived parameters such as oxygen consumption, oxygen utilization coefficient, and intrapulmonary shunt fraction.

Z-0765-2019

December 21, 2018

Class I

Product summary

Firm
Edwards Lifesciences, LLC
Event
Event 81878
Status
Terminated
Classification
Class I
Quantity
459 units
Official record key
device-enforcement:Z-0765-2019

Official wording

Reason: These Swan Ganz catheters have incorrect lumen assembly, which causes reversal of the lumens. If the lumens are reversed the clinician may note reverse PA and CVP pressure values and waveforms.

Code information: Lot Numbers: 61176367, 61176370

Distribution pattern: Worldwide Distribution

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0