Skip to content
Recall Observatory FDA recall evidence

Device product

3M(TM) Surgical Clipper by Remington(R) REF 9602 The 3M Surgical Clipper Classic, 9602, is intended to remove body hair and head hair from patients in preparation for any medical procedure requiring hair removal.

Z-0791-2019

January 14, 2019

Class II

Product summary

Firm
3M Company - Health Care Business
Event
Event 81933
Status
Terminated
Classification
Class II
Quantity
19,860 units
Official record key
device-enforcement:Z-0791-2019

Official wording

Reason: 3M has recently received reports involving the 3MTM Surgical Clipper by Remington¿, Model 9602, overheating in mobile/field EMS situations while charging from a modified sine wave power inverter.

Code information: all lots

Distribution pattern: US Distribution and countries Brazil and Canada.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0