Skip to content
Recall Observatory FDA recall evidence

Device product

Accelerate Pheno system, with Accelerate Pheno System Software 1.2.1 - 1.3.1, PSW000012B Product Usage: The Accelerate Pheno system is intended to measure signal intensity of fluorescent probes bound to nucleic acid in target and non-target organisms and to take time-lapse dark-field images of immobilized growing bacterial cells when used with Accelerate PhenoTest kits.

Z-1065-2019

March 13, 2017

Class II

Product summary

Firm
Accelerate Diagnostics Inc
Event
Event 82216
Status
Terminated
Classification
Class II
Quantity
81
Official record key
device-enforcement:Z-1065-2019

Official wording

Reason: Rare isolates of Enterobacteriaceae may generate a susceptible meropenem Minimum Inhibitory Concentration (MIC) using the affected device and a resistant result by broth microdilution. In order to mitigate any risk associated with these rare isolates, the firm has implemented an Expert Rule that suppresses the meropenem result for isolates that display a profile consistent with those that generated a false-susceptible result. The firm has responded to this risk by implementing a set of suppression rules, in a software update, to prevent incorrect results.

Code information: REF: 10401008. Serial Numbers: 1324, 1325, 1431, 1432, 1084, 1087, 1367, 1362,1363,1615, 1616, 1358, 1359, 1453, 1354, 1357, 1599, 1176, 1178, 1353, 1356, 1576, 1577, 1578, 1322, 1323, 1454, 1594, 1598, 1288, 1290, 1688, 1690, 1691, 1694, 1798, 1799, 1800, 1801, 1135, 1151, 1326, 1327, 1477, 1478, 1479, 1480, 1619, 1620, 1617, 1618, 1489, 1501, 1687, 1573, 1574, 1144, 1145, 1247, 1249 ,1264, 1277, 1265, 1277, 1278, 1381, 1150, 1286, 1148, 1149, 1254, 1260, 1397, 1542, 1458, 1517, 1231, 1238, 1282, 1284, 1389

Distribution pattern: Worldwide Distribution - US Nationwide North America and Middle East US: TX, IL, TN, GA, CA, AR, MD, VA, Washington DC, IA, PA, NY, WA, and MO. OUS (Foreign): UK, Spain, Italy, and Germany

Derived failure modes

  • Device software or design

    reason.device_software_design · v1.0.0

    software update, to prevent incorrect